• Gas Chromatograph GC UK

    Gas Chromatograph GC
    Model: GC650
    Brand: PG instrument
    Origin: UK

    Technical Specifications
    Column Oven
    Inner volume 16.25 liter (250mm W* 260mm D*250mm H) Temperature Range Ambient to 400℃ Linear Programming Rate Max: 120℃/min up to 400℃ (0.1℃ increment) Heating Rate About 2.3 minutes 50℃ to 300℃ Cooling Rate About 6 minutes 300℃ to 50℃ Run Time Automatically computed to 999.9 minutes by 0.1 minute increments Temp. Programming 30 Step

    Pneumatics (IPC Module)
    IPC pressure range 0 to 100psi IPC flow range 0 to 450mL/min Pressure resolution 0.01psi Flow resolution 0.01mL/min Support IPC Channels Up to 20, for inlets, detectors or auxiliary gases (including pressure channel) Pressure/flow programming steps up to 5 steps EPC control Inlets and detectors (carrier, split vent, make-up and combustion gases) Split ratio up to 5000:1

    Detector Modules Flow set range
    FID/NPD make-up 0-100 mL/min FID air 0-450 mL/min FID hydrogen 0-100 mL/min TCD make-up 0-130 mL/min TCD reference 0-100 mL/min FPD make-up 0-130 mL/min FPD air 0-250 mL/min FPD hydrogen 0-300 mL/min

    Inlet Modules Pressure sensor
    Accuracy ±2% full scale Repeatability ±0.05psi Temp. coefficient ±0.01psi/℃ Inlet Modules Flow sensor Accuracy ±5% depending on carrier gas Repeatability ±0.35% of set point Temp. coefficient ±0.20mL/min normalized temp. and pressure

    Data Acquisition System
    Dynamic range >2*106 Sampling Rate User selectable 10-60Hz per channel Analog signal voltage ranges ±10,5,2.5,1.25V Analog to digital conversion resolution 21 bits at all sampling rates External event I/O ports 8 Remote Start, stop control independent Data Visualization Up to chromatograms in real time, zoom/unzoom at any time; Scrutinize focused peaks, display baseline; Chromatograms overlap, background subtraction. Data Management Data export results as Text files; Dynamic data exchange to Microsoft windows applications; Various data management techniques support the time event programming (17items). Custom reports Peak number, retention time, peak name, peak area, peak height, area%, height%, peak start time, theoretical plates, capacity factor, peak resolution, peak asymmetry, regression coefficient value, correlation coefficient value, standard deviation, weight function, relative retention time, calibration equations.


    Dimensions, weight and power
    Dimensions 425mm(W)*555mm(D)*450mm(H) Weight 30Kg Power 220V, 50/60Hz,2.7kW

  • GC  UK