Autoclaves
Model: KTR-40S
Brand: ALP
Origin: Japan
Technical Specifications
Volume: 62 Litres
Chamber (dia. x depth): Ø 400 x 500 mm Usable max. press.: 0.16 Mpa
Usable max. temp.: 1250C Sterilization: 110 ~ 1250C
Sterilization system: saturated steam generated by sheathed electric heater
Main controller: Digital temperature/ time controller
Timer: 1 – 60 min. 59 sec (operates after the set temperature is reached)
Operation star: PUSH-ON start button swith
Air exhaust valve: Gravity substitution system by temperature sensing-time controlled solenoide vale
Water lack preventor: Automatic cut-off system by a sensitive thermostat and relay (Heater cut off, Buzzer & Red lamp warning)
Exhaust valve: manual needle valve to be opened only in rapid exhaust required
Drain valve: manual ball valve (inner dia.: 10 mm)
Sterilization pattern: (Tempereture/ time set) → PUSH-ON start → Sterilization → Natural cooling → Completion buzzer. RUN lamp off at safe temp. for lid open (800C)
Safety devices: Water lack preventor, Over pressure safety switch, Saftey valve
Electric safety: Breaker for over current & earth leakage, withstand voltage: 1400V, 2 sec., insulation: 100 MΩ ~ (500VDC) Safety valve operation: 0.18 Mpa Water test pressure: 0.36 Mpa
Chamber & lid: stainless steel SUS304 shiny polished, lid gasket: molded silicon rubber
Outer panel: upper surface: SUS304, main body: steel with baked enamel, caster (front two with stopper) Electricity: AC 220 V, 1 phase, 50/60Hz Power requirement: 3.6 kW (18 A) Outer size: 770W x 550D x 810H mm
Accessories:
+ Stainless wire basket Ø 380 x 200 mm, 2 pcs
+ Exhaust drain receiver bottle, 1 pc
+ Stainless drain hose, 1 pc